MAN205-0195042EN Rev. A
Aug. 2011
Page 55 of 137 pages
LimitP add
– Adding pressure to the maximum pressure
This value defines the addition to the maximum pressure, computed to the jacket
control.
For example,
when the desired temperature is 134ºC (273ºF) the maximum
temperature maintained in the jacket will be adjusted to a temperature of 137ºC (i.e.
332kPa) - 278 ºF (48 psi)
When this pressure is not enough, it can be increased by changing the values of the
LimitP add.
If the parameter value remains Ø, the maximum pressure in the jacket will be 332kPa,
however if this value will be 10, the maximum pressure will be 342kPa.
Entry code
11
Resolution
1 kPa (0.5 psi)
Minimum value
ØkPa/psia
Maximum value
30 kPa (4.35 psia)
PulsPres1
– The pressure in pulse no.1 in the pre-vacuum stage
This defines the first pulse pressure in the pre-vacuum stage. If it is defined as 100kPa,
the system will not enter steam into the chamber, however; the vacuum pump will
immediately start operating to create a vacuum.
If the value is defined over 100kPa, the system will open the steam valve to the chamber
until the desired pressure is reached.
Entry code
11
Resolution
5 kPa (1 psi)
Minimum value
100 kPa (14.5 psia)
Maximum value
300 kPa (44 psia)
PulsPres2
– The pressure in pulse 2 and onwards in the pre-vacuum stage
After the completion of pulse no.1, the system will enter steam by the defined value in
pulse press.
Entry code
11
Resolution
5 kPa (1 psi)
Minimum value
100 kPa (14.5 psia)
Maximum value
300 kPa (44 psia)
CODE 13 - PARAMETER
- The addition to the required pressure for sterilization
This parameter sets the pressure point maintained during the sterilization stage in the
following manner:
If the desired temperature is 134ºC (273°F), the required pressure (from the tables) is
305kPa (44.2 psi).
When the is Ø, the system pressure will be maintained at 305kPa (44.2
psi)during the sterilization stage. When the equals 5kPa (0.5 psi), the system
will be maintained at 310kPa (44.9 psi) in the sterilization stage. It is advised to work
with a value of 7kPa (0.5 psia).
Entry code
13
Resolution
1 kPa (0.1 psi)
Minimum value
2 kPa (0.1 psia)
Maximum value
20 kPa (4 psia)
Summary of Contents for 6671130
Page 2: ...Web site http www tuttnauer com...
Page 30: ...MAN205 0195042EN Rev A Aug 2011 Page 24 of 137 pages PIPING SYSTEM DRAWING...
Page 32: ...MAN205 0195042EN Rev A Aug 2011 Page 26 of 137 pages PNEUMATIC CONTROL SYSTEM DRAWING...
Page 84: ...MAN205 0195042EN Rev A Aug 2011 Page 78 of 137 pages This page intentionally left blank...
Page 128: ...MAN205 0195042EN Rev A Aug 2011 Page 122 of 137 pages This page intentionally left blank...
Page 129: ...MAN205 0195042EN Rev A Aug 2011 Page 123 of 137 pages XPCS MANUAL...