Page 74 of 77
Problem/
Error
Message
Message /
Symbol Description
Corrective Action
"High Temp.
(Ending)"
This message is displayed if the system
cannot reach the required temperature, in the
chamber, within 10 minutes.
Perform a new cycle.
"Heat Time
Error"
This message is displayed if the system
cannot reach the required temperature, in the
chamber, within the preset time.
Verify that the autoclave is not
overloaded.
"Heat Time
Error (Keep)"
This message is displayed if the system
cannot reach the required temperature, in the
chamber, during the optional "Keep Heat"
stage, within the preset time.
Verify that the autoclave is not
overloaded.
"Low
Pressure"
This message is displayed if Chamber
Pressure drops below the sterilization
pressure (135°C = 313 kPa ,134°C = 304
kPa ,132°C = 286 kPa 121°C = 205 kPa)
for 2 seconds during the sterilization stage.
Perform a new cycle.
"High
Pressure"
This message is displayed if Chamber
Pressure raises 4.2 psi-29 kPa above
sterilization pressure (135°C = 313 kPa
,134°C = 304 kPa ,132°C = 286 kPa 121°C
= 205 kPa) for 2 seconds during the
sterilization stage.
Perform a new cycle.
"High
Pressure
(Ending)"
This message is displayed if the system
cannot reach atmospheric pressure ± 5kPa
during the ending stage.
Perform a new cycle.
"Pressure Time
Error"
This message is displayed if the system
cannot reach the required pressure
conditions in the chamber, after preset time,
during the air removal stage.
Verify that the autoclave is not
overloaded.
"RTC Error -
Please Set
Current Date
and Time"
This message is displayed in order to set
the date and the time.
Set Current Date And Time.
If the problem persists, call the
technician.
"Time Error"
This message is displayed if the real time
clock is faulty.
Call the technician.
"Door is open
(During the
cycle)"
This message is displayed when the door is
open: During the cycle.
Close the door to perform a new
cycle.
"Canceled By
User"
This message is displayed after the
START/STOP key is pressed and cycle
aborted.
Wait until "cycle failed – canceled
by user" or "cycle end – canceled
by user" is displayed.
Perform a new cycle.
"Cycle Failed" This message and symbol are displayed if
an error occurs before sterilization cycle is
completed.
Perform a new cycle.
Summary of Contents for 2840 EL-D
Page 4: ...Page 3 of 77 14 TROUBLESHOOTING 73 15 SPARE PARTS LIST 77 16 ACCESSORIES 77...
Page 5: ...Page 4 of 77 TABLE OF CONTENT Cont DRAWINGS PAGE NO FRONT VIEW 17 REAR VIEW 18...
Page 26: ...Page 25 of 77...
Page 34: ...Page 33 of 77...
Page 71: ...Page 70 of 77...
Page 72: ...Page 71 of 77 Cap Gasket Strainer element Strainer Housing WATER STRAINER...
Page 77: ......