Page 13 of 77
3.6
Utilities
Specifications
Value
Mineral free water
See table in sec. 3.9.
Power supply
* 1 phase, 208/230VAC ±10%,
60Hz
Recommended circuit
breaker
15A
* According to the local network.
Attention:
·
A switch or circuit-breaker must be included in the building installation.
This switch or circuit-breaker shall be in close proximity to the
equipment, within easy reach of the operator; and marked as the
disconnecting device for the equipment.
·
The electrical net must be protected with a current leakage safety relay.
·
The electrical network must comply with local rules or regulations.
·
Verify that there is an easy access to the main power switch, to the water
cut-off valve and to the current leakage safety relay.
3.7
Environment Emission Information
1. The peak sound level generated by the autoclave is 65dBa with
background noise of 48 dBa.
2. The total heat per hour transmitted by the autoclave is <200Wh.
3.8
Construction
The main parts of the autoclave are made of materials as indicated below:
·
Chamber is built of stainless steel.
·
Door is made of stainless steel.
·
Trays are made of stainless steel.
·
Water reservoir is made of hard plastic material.
·
Door handle is made of hard plastic material, which is safe to touch
and thermo-insulated.
3.9
Loading Capacities
3.9.1
Erlenmeyer Flasks
Size
250 ml
300 ml
500 ml
1000 ml
Qty
8
8
4
2
3.9.2
Medium Flasks (Schott)
Size
250 ml
500 ml
1000 ml
Qty
10
8
4
Summary of Contents for 2840 EL-D
Page 4: ...Page 3 of 77 14 TROUBLESHOOTING 73 15 SPARE PARTS LIST 77 16 ACCESSORIES 77...
Page 5: ...Page 4 of 77 TABLE OF CONTENT Cont DRAWINGS PAGE NO FRONT VIEW 17 REAR VIEW 18...
Page 26: ...Page 25 of 77...
Page 34: ...Page 33 of 77...
Page 71: ...Page 70 of 77...
Page 72: ...Page 71 of 77 Cap Gasket Strainer element Strainer Housing WATER STRAINER...
Page 77: ......