Page 45 of 77
12.2
Table of parameters
The parameter default depends on the selected program. The following
table is an example to Custom 1 program.
Parameters
Description
Default Unit Range
Resolution
Pulse A Count Define the number of pulses in the first pulses group
3
#
0-10
1
Pulse A Stay
Time
Define the additional time that the vacuum pump will
continue to operate or slow/top exhaust valve will remain
open after reaching the preset pressure in the first pulses
group
10
Second 1-100
1
Pulse A Low
Pressure
Define the minimum vacuum/pressure level in the first
pulses group
130.0
kPa
5-200
1
Pulse A High
Pressure
Define the maximum vacuum/pressure level in the first
pulses group
180.0
kPa
5-200
1
Pulses B Count Define the number of pulses in the second pulses group
0
#
0-10
1
Pulse B Stay
Time
Define the additional time that the vacuum pump will
continue to operate or slow/top exhaust valve will remain
open after reaching the preset pressure in the second
pulses group
2
Second 1-100
1
Pulse B Low
Pressure
Define the minimum vacuum/pressure level in the second
pulses group
160.0
kPa
5-200
1
Pulse B High
Pressure
Define the maximum vacuum/pressure level in the second
pulses group
180.0
kPa
5-200
1
Sterilization
Temperature
Define the sterilization temperature
134.0 °C/°F 80-137 0.5°
Sterilization
Time
Define the sterilization time
4.0
Minute 0-9999
0.5
F0Mode
This parameter calculates the time that the load is exposed
to temperature according to F0 table. The printer prints
out the values before sterilization and at the end of this
cycle
0
#
0-1
1
End
Temperature
Define the temperature at the end of the cycle , the cycle
will not end before reaching the preset temperature
120
°C/°F 30-120
1
Multiple
Cycles
Define the number of times the cycle repeated
1
#
1-50
1
Multiple
Cycles Gap
Define the interval between repeated cycles in case that
Multiple Cycles parameter is defined
2
Minute 1-60
1
Print Rate All
Defines the printing rate during the cycle except
sterilization stage
3
Minute 1-30
1
Print Rate
Sterilization
Defines the printing rate during the sterilization stage
1
Minute 1-30
1
Screen Saver
Defines the interval from the last use of the Keypad until
activating the screen saver
90
Minute 0-600
1
Summary of Contents for 2840 EL-D
Page 4: ...Page 3 of 77 14 TROUBLESHOOTING 73 15 SPARE PARTS LIST 77 16 ACCESSORIES 77...
Page 5: ...Page 4 of 77 TABLE OF CONTENT Cont DRAWINGS PAGE NO FRONT VIEW 17 REAR VIEW 18...
Page 26: ...Page 25 of 77...
Page 34: ...Page 33 of 77...
Page 71: ...Page 70 of 77...
Page 72: ...Page 71 of 77 Cap Gasket Strainer element Strainer Housing WATER STRAINER...
Page 77: ......