Page 75 of 77
Problem/
Error
Message
Message /
Symbol Description
Corrective Action
"Air Error"
This message is displayed at the end of
the cycle If the autoclave does not reach
the atmospheric pressure after 10 minutes.
Wait until the autoclave reaches
the atmospheric pressure and
perform a new cycle.
"Periodical
check time
exceeded -
Please call for
service"
The periodical maintenance time has
passed.
Call for service.
"Mineral free
water
reservoir
empty"
This message is displayed if the water
level electrode does not sense water.
Fill the mineral free water
reservoir.
"Cycle counter
exceeded -
Please call for
service"
Number of cycles, since last periodical
maintenance, exceeded the preset number
as defined by "cycle counter" parameter.
Call for service.
"Power
Down"
This message is displayed if power down
has occurred during the cycle. (this
message will print out in the printer after
the autoclave turns on).
Turn on the autoclave and wait
until the autoclave is ready
(reaches the safe condition) and
perform a new cycle.
"No Water"
This message is displayed if the electrode
in the chamber did not sense water within
the preset time.
1. check and fix the mineral free
water supply.
2.check and clean the water inlet
filter.
3. Clean the water level electrode.
Summary of Contents for 2840 EL-D
Page 4: ...Page 3 of 77 14 TROUBLESHOOTING 73 15 SPARE PARTS LIST 77 16 ACCESSORIES 77...
Page 5: ...Page 4 of 77 TABLE OF CONTENT Cont DRAWINGS PAGE NO FRONT VIEW 17 REAR VIEW 18...
Page 26: ...Page 25 of 77...
Page 34: ...Page 33 of 77...
Page 71: ...Page 70 of 77...
Page 72: ...Page 71 of 77 Cap Gasket Strainer element Strainer Housing WATER STRAINER...
Page 77: ......