Infrared Camera – Operating Manual

 

B -

 

6

Powering On or Off

The power indicator shows that the camera is on and 
has suffi cient battery power.
1.  Hold  the  camera  as 
    shown, place your thumb 
    above the keypad and rest 
    your  forefi nger  gently  on 
    the trigger. 

2.  Press and hold the power 
    switch for 3 seconds. The 
    power  indicator  lights  up 
    green.

3.  Shortly afterwards, a start-
    up image appears on the 
    screen.

4.  To power off press the power key and hold for 3 se-
    conds. The power indicator goes off.

Information on the LCD Monitor

The LCD monitor has a fi eld of vision of 100%.
The following view provides you with information on 
individual data:

Information regarding the operating indicator

The  operating  indicator  shows  which  operating  status 
the camera is in.

Menu . . . . . . . . . . . . . . . . . . . . .shows the menu mode

Zero . . . . . . . . . . . . . . . . . . . . . . . shows the non-menu 
                                            No analysis tools is selected

1 - 4 . . . . . . . . . . . . . . . shows the current analysis tool 
                                            is sport 1 or spot 2 or spot 3

Cap  . . . . . . . . . . . . . . . .  shows that the tool has been 
                                      selected  to  automatically  track
                                    the measuring point (auto tracking)

Isot . . . . . . . . . . . . . . . . . . . . shows the current analysis 
                                                     tool is isotherm analysis

E. . . . . . . . . . . . . . . . . . . . . . . . . . . . .  current emissivity

      . . . . . . . . . . . . . . . . . . . . . . . . . . . . SD card inserted

. . . . . . . . . . . . . . . Bluetooth headset has been installed

you  will  need  to  enter  the  zero  mode 

before  you  start  measuring  with  your 

camera. In order to do so press the Can-

cel key repeatedly until the required zero 

mode appears in the display.

m

Reading

Status of the camera

Live/
Freeze

Operating indicator

Current Emissivity

Battery Status

time

Upper limit of 
colour scale

Lower limit of 
colour scale 

Colour scale

NULL E=1.00

ZERO

Summary of Contents for POWERSHRED

Page 1: ...TROTEC GmbH Co KG Grebbener Stra e 7 D 52525 Heinsberg Tel 49 2452 962 400 Fax 49 2452 962 200 www trotec de E Mail info trotec de TRT EC060 HS 003 GB EC Series EN Infrared Camera Operating Manual B 1...

Page 2: ...n B 17 06 Recording and Deleting B 17 Opening Images B 17 Selecting Images B 17 Selecting the Name of the Current Folder B 18 Voice Annotations B 18 Deleting Images B 18 07 Reading Out Images B 19 Rea...

Page 3: ...turn off the power of the camera remove thebattery from the camera or unplug the power cord from the power outlet Check to see that no more smoke or fumes are coming out of the camera Stop operating i...

Page 4: ...en the IR camera is not in use to avoid fires and other hazards Non stop operation use over a long period of time may cause the device to overheat and deform resul ting in fire Do not use the battery...

Page 5: ...s Locking Ring Trigger Multi functional Docking Station Rear View KeyPad LCD Monitor USB Terminal Video Output Terminal Battery SD Card Cover Controls Multi function Docking Station Auto Adjust Key Po...

Page 6: ...le charger to charge your battery When the charging cycle has been completed discon nect the charger from the mains supply and remove the battery Your camera is equipped with a lithium ion battery whi...

Page 7: ...the operating indicator The operating indicator shows which operating status the camera is in Menu shows the menu mode Zero shows the non menu No analysis tools is selected 1 4 shows the current anal...

Page 8: ...settings This menu item can be used to change the style of the menu system 1 Check that the IR camera is in zero mode 2 Press the MENU EN TER key then press the up s or down t arrow on the omni select...

Page 9: ...ng target is in the centre of the display The LCD monitor will switch itself off when closed Selecting Menus and Settings You can select the settings by pressing the MENU ENTER key 1 Press the MENU EN...

Page 10: ...ill not be deleted when you reset the menu and button operation settings to default 05 MEASURING Manual Focusing 1 Check to ensure that the IR camera is in null mode 2 Point the camera at the object y...

Page 11: ...y and the arrow keys in order to adjust the thermal image so that it covers the visual image Moving the DuoVision area Move upwards C UPs Move downwards C DOWNt Move left C LEFT Move right C RIGHT WAR...

Page 12: ...NU ENTER key to save the changes or press the C key to close the window without saving any new settings Information regarding image settings Auto adjust Sets the function of the A key There are three...

Page 13: ...close the window when this step is completed Freezing Activating an Image You can freeze or activate a thermal image by pres sing the S key on the selector 1 Check that the IR came ra is in zero mode...

Page 14: ...down t arrow on the omni selector to select the menu item Analysis then press the MENU ENTER key 3 Press the up s and down t arrow on the omni selector to select Setup then press the MENU ENTER key 4...

Page 15: ...en 1 and 100 and describes the proportion in between 1 100 and 100 100 of the area of the LCD eg the isotherm range is between 35 C and 40 C and the Iso Alarm is 100 In this case the alarm is triggere...

Page 16: ...andAverage temperature of the selected area Press the left or right key to select between the Max Min or Average temperature of the selected area When Area 5 is selected the Max Min and Ave rage temp...

Page 17: ...therm range by changing the lower isotherm level IL The letters IL and IH appear in the bottom right hand corner IH is the upper threshold and IL the lower threshold of the isotherm range Removing Ana...

Page 18: ...down arrow on the omni selec tor to select menu item Others and then press the MENU ENTER key Trigger Information Save Press the trigger to save an image Laser on Press the trigger to activate the la...

Page 19: ...luetooth headset optional and open an image 2 Press the MENU ENTER key then press the up or down arrow on the omni selector to select the File menu and confirm 3 Press the s up or down t arrow on the...

Page 20: ...ture when the camera is used in continuous operation Ple ase ensure that you only use original Trotec mains adapters to avoid any damage from occurring to your camera Connecting to a monitor The camer...

Page 21: ...stall programs 1 After a few moments the following dia logue will appear 2 Select No notthistime then click Next 3 Select Installfromalist or location Advanced then click Next 4 Select Include this lo...

Page 22: ...etooth headset first Press and hold the power button A for about 5 seconds The power indicator starts to blink in red and then blue The headset is in pairing mode after 2 seconds ThebLEDblinksmoreslow...

Page 23: ...l Save the images onto your computer and delete them from the SD card Internal memory not formatted correctly Format the memory in FAT32 format Battery power exhausted Battery capacity reduced because...

Page 24: ...00 1400 0 29 Melting mild steel 1600 1800 0 28 Liquid steel 1500 1650 0 42 0 53 Pure lead 1515 1680 0 42 0 45 Silver Polished Silver 100 0 05 Material Temperature C Emissivity approx Nickel Nickel chr...

Page 25: ...na 25 600 0 55 Mild alumina 25 600 0 10 0 20 Strong alumina 25 600 0 30 0 40 Copper Cuprous oxide 800 1100 0 16 0 13 Material Temperature C Emissivity approx Non metallic materials Mica 0 94 0 95 Flam...

Page 26: ...res Image freeze Live or freeze File function Removable SD card up to 2 GB Measurement Temperature range 20 C 250 C Accuracy 2 C or 2 Analysis Functions Spots 2 10 spots Temperature tracking Tracks th...

Page 27: ...or 8 11V DC supply to charge the battery inside BWT 2 5 hours Environmental specification Operational temperature 15 C 50 C Storage temperature 40 C 70 C Relative humidity 10 95 Interfaces Mechanism w...

Page 28: ...TROTEC GmbH Co KG Grebbener Stra e 7 D 52525 Heinsberg Tel 49 2452 962 400 Fax 49 2452 962 200 www trotec de E mail info trotec de...

Reviews: