Treedental TR-KID-7 Instructions For Use Manual Download Page 3

-1-

I. Safety Information

Be sure to read through this operating manual to obtain a correct using of this unit. In case of any special requirement, please

contact the dealer of our company.

Warning:

1. Please neither put the appliance in moist environment nor open the tank body for the sake of reducing the possibility of fire
and electric shock. In case of maintenance, please refer to professional maintenance man.
2. In case of any smoke or other abnormal circumstance which may result in fire or other conditions, please unplug the power
connection and contact the local dealer or the after-sale service department.
3. In case water enters into the equipment, which may result in creepage, please unplug the power connection and contact the
local dealer or the after-sale service department.
4. In case of any undefined fault of the equipment, please unplug the power connection and contact the local dealer.
5. Please don’t disassemble or refit this product, otherwise, some accessories and components may be damaged or result in
electric shock.
6. The power line shall not be damaged. In case the power line is damaged, please disconnect the power supply. Don’t touch the
power supply with hand directly. Please contact the local dealer to change the damaged power line.
7. The product is adaptable to the voltage of a.c.230V/50Hz. In case of work under other voltage, dangers may occur.
8. Please care the weather change. In case of thunder, please disconnect the power supply.
9. Please don’t touch the power plug with wet hands, which may result in electric shock.
10. Please don’t use insecticide, thinner and chemical liquid to clean the equipment, which may peel or rot its surface.
11. Please don’t remanufacture the machine. In case of any damage consequently, it is beyond any maintenance and repair
guarantee. No modification of this equipment is allowed.
12. Contraindication: NO
13. Warning: To avoid risk of electric shock, this equipment must only be connected to a supply with protective earth

14.

The power supply cord can’t be replaced by service personnel.

15. The mains plug is considered as isolation device from supply mains.
16. Do not to position the equipment to make it difficult to operate this disconnection device.

To users:

This product has undergone the test, which conforms to the requirement of EU’s 93/42/EEC 2007/47/EEC

instruction. No hydrargyrum segregator has been installed within the product; therefore, its drainage system cannot filter
solid particles and debris, or amalgam alloy. In order to avoid such pollution to the drainage system, the user shall install
a hydrargyrum segregator additionally to avoid the drainage pollution.

Advice to arrange waste amalgam, dentist should be collected waste amalgam monthly, and should be according with

environmental regulations, and supply them to the cerfitied waste management company.

If you want to know more related information, please contact us.

Please record the product code of this appliance here for the convenience of inquiry in the

future.

Summary of Contents for TR-KID-7

Page 1: ...rapy unit produced by our company Please read this manual carefully for making best use of all functions of this product Please keep this manual for reference in case of problem occurred to the dental...

Page 2: ...ntrolling 18 2 Emergency stop switch 20 3 Assistant tray contolling panel 20 4 Water tank 21 5 Foot controller 21 6 Valve of handpiece 22 7 Operating light 22 8 X ray film view 23 9 3 way syringe and...

Page 3: ...er change In case of thunder please disconnect the power supply 9 Please don t touch the power plug with wet hands which may result in electric shock 10 Please don t use insecticide thinner and chemic...

Page 4: ...nfiguration in which it will be used Guidance and manufacture s declaration electromagnetic emission The TR KID 7 is intended for use in the electromagnetic environment specified below The customer of...

Page 5: ...arth 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply...

Page 6: ...mpliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 V m NOTE 1 At 80 MHz a...

Page 7: ...W Separation distance according to frequency of transmitter m 150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 F...

Page 8: ...N for part of equipment Dental patient chair tilt backward OFF for part of the equipment Dental patient chair tilt forward Emergency Stop Spittoon position Dental operating light Bowl flush X ray Film...

Page 9: ...agile handle with care Keep away from sunlight Protect from heat and radioactive sources Upward Prohibited stacking of 2 Stacking limit by mass of 350 kg Symbol for the marking of electrical and elect...

Page 10: ...ating stool doesn t belong to the basic structure of the dental multipurpose therapy unit and the guarantee period of the operating stool is three months The cold light lamp PU cushion real leather cu...

Page 11: ...requirement for cleanness should be met The air compressor should be provided by the user and compressed air must be subject filtration with a filter pore size 50 m 2 Water supply water pressure 0 20...

Page 12: ...ad rack extension range 120mm The biggest backward angle of the seat cushion 12 Temperature of the constant temperature water heater 40 5 Work space length 4000mm width 2600mm height 2000mm Handpiece...

Page 13: ...11 Fuse label...

Page 14: ...21 22 26 23 25 29 USB Motor Input 24V AC Limited position swtich procedure control box USB USB operating light film viewer 24V water heater and temperature controller AC 24V output 1 2 3 4 5 6 water...

Page 15: ...e touching from any objects during the dental units in the working the installation space should be required the following items a The length no shorter than4000mm b The width no shorter than 2600mm c...

Page 16: ...0mm then connect the cable plugs from patient chair please connect them according to correct numbers to the cables from spittoon unit fix the spittoon unit to patient chair by 3 pieces M10 he socket 3...

Page 17: ...ug Operating Lamp figure 2 7 Power unit Connect the water supply tubing air supply tubing and drainage tubing to water source air source and drainage respectively During the connecting make sure that...

Page 18: ...insert of the nozzle is not finished The meter ring should be pressed again to make the nozzle inserted till the meter ring recoils b hand piece turn on the water source and air source before install...

Page 19: ...st be 52 mm PVC pipe The pipe interfaces should be about 40mm from ground Connection area for electricity should be properly grounded and within the installation area 4 Pipeline installation Before li...

Page 20: ...er in the front cover The normal operating air pressure of the dental unit is 0 55Mpa If any error occurs please adjust the knob in the air filtrating and reducing valve to keep the above mentioned nu...

Page 21: ...ot water Press the constant temperature heater will start working to heart the water to 40 in case of water temperature decrease the heater will work automatically the indicator light flashes in the c...

Page 22: ...water supply if tap water changes please reset the time a Memory position set of the patient chair Step 1 press SET key the indicator light turns on Step 2 press key or key press one of them or neithe...

Page 23: ...fiers direction release the switch which will then return to its original position automatically and then the patient chair will stop UP the patient chair upward DO the patient chair downward BD backr...

Page 24: ...ir adjustment valve Handpiece 2 air adjustment valve Handpiece 3 adjustment valve Figure 1 Figure 2 Figure 3 a 1 2 3 in Figure 1 and Figure 2 correspond to the handpiece water and air amount of 1 2 3...

Page 25: ...ment instructions To avoid the risk of biocompatibility please use the saliva ejector conform with ISO 10993 Suction low is water suction Connection through water source is required and minimum water...

Page 26: ...lear the filter element of water fliter by mild detergent at least every 6 months the specification of filter core is less than 50 m if there is cracking or breakage in the water filter or filter cove...

Page 27: ...EL 9 Annex Spare parts list for repairing by Maintenance staff 10 Dental handpiece and 3 way syringe tip need to be sterilized before use It is recommended to use 135 high temperature sterilization 15...

Page 28: ...ce drips water even not under use 1 Malfunction of the air distribution valve 2 The air controller foot switch doesn t return to the original position 1 Settle or change a new pipe 2 Change a new elec...

Page 29: ...rum is not well connected 2 The malfunction of the keys on the main control panel 3 The control box is burned such as the fuse is burned 4 The motor is broken 5 The limit switch is blocked 1 Re plug t...

Page 30: ...etic valve DC24v water heater water heater temperature controller DC24V output strong pumpping electromagnetic valve Weak pumpping electromagnetic valve strong pumpping switch Weak pumpping switch Neg...

Page 31: ...n instrument tray electromagnetism valve high power suction water oneway valve P A air adjustment valve water transfer switch tank 2 1 air air brake setting instrument tray air brake switch 3 4 handpi...

Page 32: ...in pipe 3 way syringe air brake valve tank high speed heedpiec 1 high speed heedpiec 2 high speed heedpiec 3 water regutating valve holder valve chip blowing valve air adjustment valve left foot switc...

Page 33: ...staff 2 electric motor Professional maintenance staff 14 air meter Professional maintenance staff 3 tubing Professional maintenance staff 15 valves Professional maintenance staff 4 main control keyboa...

Page 34: ...with care and moisture proof 2 Transportation and storage environment temperature 5 40 relativity 80 pressure altitude 2000m Installed by Telephone Our company reserves the right to change product des...