III. Matters of Concernment:
This product can only be used for dental treatment, therapy and operation.
The children are forbidden to play near the equipment in order to avoid accidents while touching the equipment.
Only use the fuse and accessories recognized by our company can be used.
This dental unit should not come into contact with any object in its operational range of movement .
After using this dental unit r if this dental unit is not os not to be use the patient chair be returned to the lowest position and,
water air , and the power supply be cut off.
Make sure the Headrest is locked before using the patient chair.
Don’t use the product in high-temperature, high-moisture, vibration, flammable or unbalanced environment.
The patient chair is intermittent operation equipment. ON/2,MIN, OFF/ 18 MIN.
Read the use instructions before using the handpiece.
Read the use instructions before using the curing light machine and the scaler
Connect the water supply before using the water heater in order to avoid short circuit with of dry heating.
Contact professional Service personnel for unit repairing or maintenance, disconnect the water, air and power supply before
the repair.
Disconnect the water, air and power supply while not using the appliance for long time.
It is advised that The expected service life of the dental unit is 5 years for this dental unit. The excessive using may result in
aging pipes, malfunctions or even accidents. Please contact qualified medical appliance recycle unit to eliminate or change
new equipment in time.
The operating stool doesn’t belong to the basic structure of the dental multipurpose therapy unit and the guarantee period of
the operating stool is three months. The cold light lamp, PU cushion, real leather cushion, glass parts and ceramic parts
within the machine are damageable parts and their guarantee period is all three months. The product’s quality problems due
to manual reasons are beyond the guarantee area.
The product is forbidden to be connected with the power supply for more than 24 hours. It is the intermittent equipment.
Please disconnect the water, gas and power supply before duty off
Patient position requirement: :operator regulate the best position during the treatment , patient lie on the dental chair ,
operator is closed to patient’s head to operate
Safety terminate operation of equipment: Press emergency safety switch at the rear side of the dental chair, all the operating
would be stopped
When replacing the fuse, unplug the power plug. The power cord is welded. If you need to change the power cord and fuses
please contact with after sales stuff.
Attention! Only instruments which include anti-retraction devices are to be used together with the dental unit
The temperature of the applied part may up to 41
℃
due to continuous treatment. Extended skin contact may cause minor
burns.
Mechanical safety
Give adequate attention to patients and this device when it is occupied, with a view to avoiding striking or squeezing the
patient. Turn off system power immediately if this device moves accidentally and will not stop moving
IV. Transportation and Storage
1. Transportation and storage conditions: be upright; handle with care; and moisture-proof.
2. Transportation and storage environment: temperature 5-40
℃
, relativity≤80%,pressure altitud
e≤2000m
Summary of Contents for TR-KID-7