![Treedental TR-KID-7 Instructions For Use Manual Download Page 29](http://html1.mh-extra.com/html/treedental/tr-kid-7/tr-kid-7_instructions-for-use-manual_1155111029.webp)
-27-
4. Readjust the position of the rack
hanging valve.
5. Adjust the water and air pressure.
The
motor
doesn’t work
1. Check whether the thrum is not well
connected.
2. The malfunction of the keys on the main
control panel.
3. The control box is burned, such as the fuse is
burned.
4. The motor is broken,
5. The limit switch is blocked.
1. Re-plug the thrum.
2. Change the keystroke.
3. Change a new main control box.
4. Change the motor of the same type.
5. Adjust or change the limit switch.