-25-
6. Please clear the suction fliter by water every day
after using the dental unit. ( the filter core is less
than 2 mm) (as figure 6)
Please refer to the Saliva ejectors tags in the removal
process.
(as figure 7)
(Warning: Please deal with the waste according to local
laws and regulations!)
Accessories such as the need to replace the contact with
the sales staff
as figure 7
7.please read the handpiece user’s manual about its
maintenance and disinfecting before using.
8.Maintenance staff qualifications: ANYE qualify
authorized dealer taking charge the after sale
service.
The MANUFACTURER will make available on request
circuit
diagrams,
component
part
lists,
descriptions, calibration instructions, or other
information that will assist SERVICE PERSONNEL to
repair those parts of ME EQUIPMENT that are designated
by
the
MANUFACTURER
as
repairable
by
SERVICE
PERSONNEL
9.Annex
Ⅱ Spare
parts
list
for
repairing
by
Maintenance staff
10.
Dental handpiece and 3-way syringe tip need to be
sterilized before use, It is recommended to use 13
5 ℃
high
temperature sterilization,15min.
11. The seat surface and the outer surface of the dental chair
should be cleaned regularly. The surface can be used for
cleaning and disinfection of medical alcohol.
12. If cleaning and disinfection as required for the dental
tubing, please loaded harmless neutral disinfectant in the
water tank to using the disinfectant system cleaned and
disinfected. If you need turnover the dental tubing, Please
contact our after sales serivce.
XIII.Guide
Troubleshooting
Trouble
Reason analysis
Troubleshooting
Remark
The equipment
doesn’t work
1. The power supply switch is not connected.
2. The fuse of the power supply switch is burned.
3. The fuse of the equipment is broken.
4. The general switch of the equipment is not
open.
1. Connect the power supply switch.
2. Change the fuse.
3. Change
the
fuse
within
the
equipment.
4. Open the general switch.
7. Please clear the suction fliter by water every day
after using the dental unit. ( the filter core is less
than 2 mm) (as figure 6)
Please refer to the Saliva ejectors tags in the removal
process.
(as figure 7)
(Warning: Please deal with the waste according to local
laws and regulations!)
Accessories such as the need to replace the contact with
the sales staff
8.please read the handpiece user’s manual about its
maintenance and disinfecting before using.
9.Maintenance staff qualifications: ANYE qualify
authorized dealer taking charge the after sale
service.
The MANUFACTURER will make available on request
circuit
diagrams,
component
part
lists,
descriptions, calibration instructions, or other
information that will assist SERVICE PERSONNEL to
repair those parts of ME EQUIPMENT that are designated
by
the
MANUFACTURER
as
repairable
by
SERVICE
PERSONNEL
10.Annex Ⅱ Spare parts list for repairing by
Maintenance staff
11.
Dental handpiece and 3-way syringe tip need to be
sterilized before use, It is recommended to use 13
5 ℃
high
temperature sterilization,15min.
12. The seat surface and the outer surface of the dental chair
should be cleaned regularly. The surface can be used for
cleaning and disinfection of medical alcohol.
13. If cleaning and disinfection as required for the dental
tubing, please loaded harmless neutral disinfectant in the
water tank to using the disinfectant system cleaned and
disinfected. If you need turnover the dental tubing, Please
contact our after sales serivce.
VIII.
Guide
Troubleshooting
Trouble
The equipment
doesn’t work
Saliva
ejectors
figure 6