Ⅹ Adjustment
Turn off the main (water, air and power)switch
( as in the figure 1)
Note: after using the dental unit, it is better to turn off
the main switch, it can be protected the spare parts,
lengthends service life, reduce troubleshooting.
When buzzer gives out a sound “du”, the power lights on
the tool tray and the aid frame turn on. And the equipment
carries on the self-check. (as shown in Figure )
- indicating light
check the connection of water source, make sure there isn’t any leakage.
check the connection of air source, make sure there isn’t any leakage.
open air source to examine the indicated number of air
pressure meter in the air reducer in the front cover. The
normal operating air pressure of the dental unit is 0.55Mpa.
If any error occurs, please adjust the knob in the air filtrating
and reducing valve to keep the above-mentioned number.
Other information see
Page 24
ⅩⅠ
Using instructions
Please make sure the water, air and power supply have been open while using the product.
All the functions set can be saved automatically in case of power off.
When the power supply resumes after the power off, there isn’tany on safety issue, except the going operation should be shutted
temperartty. If to continue the operation, please restart the unit
1.
Instructions tray panel controlling
○
1 Position of the patient chair
Press each key for long time, then the patient chair shall move according to the identifiers’ direction. Release the key, then
the move stops.
——“
”: the patient chair upward;