![Treedental TR-KID-7 Instructions For Use Manual Download Page 23](http://html1.mh-extra.com/html/treedental/tr-kid-7/tr-kid-7_instructions-for-use-manual_1155111023.webp)
-21-
4. water tank
The water stored in the water tank is only used for the instruments on the
instrument tray; when selecting and using the water tank to supply water, the following
procedures must be followed:
water tank of disinfectant can be installed on the headpiece line for disinfection,
because the disinfectant may remain in the pipeline
,
please choose the disinfectant to
local regulations, it's non-hazardous to humans
(alcohol)
.
Step 1: Press the “
” switch to disconnect the water supply;
Step2: Flip the switch “
” to “open”, the water source for the instruments on the instrument tray shall be changed into the water tank.
Notes:
If the tap water is used for the instruments on the instrument tray, please make sure the switch “
” is flipped to the “open” position.
When the switch “
” is flipped to the “Closed” position from the "open" position, air may come out; this is a normal phenomenon.
the max negative pressure is 0.25 MPa
5.Foot controller
○
1
Position of the patient chair
Pull the switch to the indicated four positions and the patient chair will move
according to the identifiers’ direction; release the switch which will then return
to its original position automatically, and then the patient chair will stop.
——
“UP”: the patient chair upward;
——
“DO”: the patient chair downward;
——
“BD”: backrest backward of the patient chair;
——
“BU”: backrest forward of the patient chair.
○
2
Gargling water supply (SUPPLY)
Step on the key lightly to supply the gargling water until to the pre-set time; step on the key again to stop water supply in the course of water supply.
○
3
Cuspidor water rush (RUSH)
Step on the key lightly to rush the cuspidor for three minutes until its stop; step on the key again to stop water rushing immediately in the course of
rushing.
○
4
Chip blow (BLOW)
Step on the key to dry the residual water in the handpiece pipe when the handpiece completes its work.
Note:
Please adjust by yourself as to the air amount for chip blowing. (Please refer to “the instructions of handpiece water and air amount
adjustment”)
○
5
Handpiece pedal
“W” on the left pedal indicates the water supply control switch of the handpiece; “A” on the right pedal indicates the air supply control switch of
the handpiece. The methods are shown as bellow:
——
In case of stepping on the left pedal only, only the water supply;
——
In case of stepping on the right pedal only, only the air supply;
2
5
4
3
1
Summary of Contents for TR-KID-7
Page 13: ...11 Fuse label...