APPENDIX C: ERROR MESSAGES
C.1 ERROR
MESSAGES
If the scale encounters an error condition, it will display a message alerting the operator. A
description of each display follows:
C.1.1 OPERATOR
ERRORS
Message
Explanation
❐❐❐❐❐❐
Indicates that the weighing capacity of the selected scale has been exceeded.
Add
Indicates that there is not enough internal resolution to calculate the APW of
an item. This means that the items you are counting are too light for the
scale to process at all.
OL
The result is too large to display. You are most likely to see this while piece
counting on a high capacity remote scale.
Err 9
Span calibration value has been lost. Re-calibrate scale.
Err 14
The keyed-in value for APW is equal to zero or exceeds the capacity of the
scale. Key-in a valid APW value.
tErr 6
Indicates that the tare weight value you are entering is not rounded to the
nearest scale division. For example, you cannot enter 0.01 lb for a 0.05 lb
increment scale.
tErr 1
Indicates that the tare weight value or APW value you are entering exceeds
the capacity of the scale.
C.1.2 CALIBRATION
ERRORS
Message Explanation
Err 0
Indicates that the test weight value you have chosen
exceeds the scale’s capacity.
Err 1
Indicates that the test weight value you have chosen is
less than 1% of the scale’s capacity.
Err 2
Occurs when you do not place the test weight on the
scale during calibration. Can also indicate that there is
an internal error in the scale.
Page 32