13-4
October 2000 © TOSHIBA TEC
MR-3011/3012 SERVICE
2)
Abnormality of motor
3)
Communication error
YES
YES
NO
YES
NO
NO
Start
Load on motor shaft normal?
Power supplied to
connector of motor?
Excitation pattern signal
supplied to connector
of motor?
Check or replace motor
Adjust mechanical load and drive system
Remove foreign object
Check or replace the PC board
Check or replace the PC board
YES
NO
YES
NO
Start
Interface cable normal?
Replace the interface cable
Communication recovered?
Repair completed
Check or replace the PC board
Check or replace the scanner
control PC board