5.2.8
Measured values display and operating procedures
5.2.9
Self-diagnosis data display operating procedures
MEASURED VALUES *
p PHASE:
123.56 N=0
T TEMP:
30.00
°
C
X DENSITY: 2.34 %TS
Move the cursor to "2" with [
→
] key, and press [SET] key to
select [ 2:MEASURED VALUES]
1
:
READ PARAMETERS
2
:
MEASURED VALUES
3
:
SELF-DIAGNOSIS
It is possible to verify the measured phase value (and
the angle rotation N), the measured fluid temperature
value, as well as the density calculated based on them.
[*] in the upper right corner will blink synchronously with
the data updating thus making it possible to verify that
the data updating is definitely taking place.
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
SL:MICROWAVE SIG.LVL
DATA :
−
53 dBm
[ESC] RETURN
Move the cursor to "3" with [
→
] key,
and press [SET] key to select [3:SELF-DIAGNOSIS]
ST:STATUS :GOOD
[ESC] RETURN
Whether the phase m easurement operation is
normally functioning or not can be verified.
The micro wave receive signal level can be
verified. Normally, the level is -45 to -75. (dBm)
1 :READ PARAMETERS
2 :MEASURED VALUES
3 :SELF-DIAGNOSIS
[DN]
(Pre. menu)
[UP]
(Next menu)
[UP]
(Next menu)
Summary of Contents for Density (Consistency) Meter LQ500
Page 1: ...6 F 8 A 0 5 2 1 OPERATION MANUAL DENSITY METER TYPE LQ500 ...
Page 2: ......
Page 51: ... 49 6 F 8 A 0 5 2 1 ...
Page 61: ... 59 6 F 8 A 0 5 2 1 ...
Page 93: ... 91 6 F 8 A 0 5 2 1 mode from the setting mode to resume the normal measurement ...
Page 113: ......