Setting menu display and operating procedures
When [2: SETTING MENU] is selected, the density output
signal and the density display are hold in the simulated
values that have been set. This warning screen will
appear once before getting into [2: SETTING MENU].
After making sure that there is no problem, press the [
→
]
key to get into [2: SETTING MENU].
1
:
MONITORING MENU
2
:
SETTING MENU
3
:
MEASURING MODE
9:LINEARIZ/CNDUCTVTY
10:ADDITIVES CORRECT
11:OTHERS
Move the cursor to the menu number with
[
→
] key, and press [SET] key.
Press the [
→
] key to get into [2: SETTING MENU].
1
:
READ PARAMETERS
2
:
MEASURED VALUES
3
:
SELF-DIAGNOSIS
Move the cursor to "2" with [
→
] key, and press [SET] key.
Test output will be
valid.
[
→
] : CONTINUE
[ESC]: CANCEL
5:SET PARAMETERS
6:ZERO CALIBRATION
7:SPAN CALIBRATION
8:ANGLE ROTATION
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
[ESC]
(Previous
Menu)
[DN]
(Pre. menu)
[UP]
(Next menu)
[DN]
(Pre. menu)
[UP]
(Next menu)
To get into [2: SETTING MENU], it is necessary to further
enter the password "8000". In the initial condition, th e cursor
is on the forth digit. Therefore, press [DN] key 4 times to set to
"8" and press the [SET] key to get into [2: SETTING MENU]. An
incorrect password will cause the following message to
appear. Return to the screen a t left by pressing any key and
then enter the correct password once again.
Input password
PASSWORD
:0000
[SET]SET,[ESC]CANCEL
PASSWORD ERROR
PUSH ANY KEY.
[ESC]
(Previous
Menu)
Summary of Contents for Density (Consistency) Meter LQ500
Page 1: ...6 F 8 A 0 5 2 1 OPERATION MANUAL DENSITY METER TYPE LQ500 ...
Page 2: ......
Page 51: ... 49 6 F 8 A 0 5 2 1 ...
Page 61: ... 59 6 F 8 A 0 5 2 1 ...
Page 93: ... 91 6 F 8 A 0 5 2 1 mode from the setting mode to resume the normal measurement ...
Page 113: ......