No. 2B771-004EN*M
9-11
NOTE:
5. Do not connect a USB device such as a flash drive to any of the external
USB connectors (the system is provided with five USB connectors) when
the system is in the Standby status. Doing so may result in the system not
starting up normally from the Standby status.
9.4.2 Recovery from the Standby status
CAUTION:
1. Do not disconnect the power plug from the outlet while the system is
resuming after being paused. The system may malfunction.
2. If either of the following phenomena occurs, press and hold down
for 5 seconds or more to turn OFF the power of the system.
The startup screen is not displayed after waiting for 30 seconds.
The patient registration screen is not displayed after waiting for
10 minutes.
If the power is not turned ON after holding down
for 5 seconds or
more, turn OFF the main power on the power panel.
Do not turn OFF the power in this manner during normal operation.
Doing so may cause the system to malfunction.
NOTE:
1. When resuming the system from pause status, confirm that the
[STANDBY]
LED is blinking and then press
.
2. Do not operate the operation panel switches or keyboard while the system
is recovering from Standby status. The system may not start up normally.
3. Normally, the ultrasound image is displayed approximately 1 or 2 minutes
after the power switch is pressed in the standby status; however, the actual
time required varies depending on the system status.
Summary of Contents for Aplio 500 TUS-A500
Page 41: ...No 2B771 004EN M 6 2 Systems with the main power switch on the side ...
Page 63: ...No 2B771 004EN M 9 3 Systems with the main power switch on the side ...
Page 92: ...No 2B771 004EN M 13 4 PIMS page 2 2 Refer to section 19 Storing Image Data ...
Page 163: ...No 2B771 004EN M 18 5 18 2 Needle Mark Display and Angle Change Procedures ...
Page 231: ...No 2B771 004EN M A 1 ...