20.4 Preventive Maintenance Performed by Service Personnel
To ensure safety and maintain optimal system performance, the following checks must
be performed by service personnel with the required training and expertise.
Contact your TOSHIBA service representative.
Check contents
Check items
Check interval
Cleaning
Transducer, transducer holder
Main body, peripheral units
Air filters
1 year
Electrical safety
Protective conductor resistance
Leakage current
Patient auxiliary current
1 year
Mechanical safety
Caster/monitor/panel operations,
mounting sections
Peripheral units
Appearance of the transducer
1 year
Checking/recording
images
Images in each mode
Images acquired using a standard
transducer
1 year
20.5 Periodically Replaced Parts and Consumable Parts
The parts used in this equipment have a long service life. However, to maintain the
safety, specifications, and performance of the system, periodic replacement of some
parts (batteries for system power backup etc.) is required (replacement interval: 1 year).
Special techniques are required for replacement of periodically replaced parts. Contact
your TOSHIBA service representative.
Note that the power cable, casters, and the magnetic sensor securing adapter (an
accessory of the Function unit) are consumable parts.
20.6 Checks During Storage
Perform preventive maintenance during storage of the system.
*
Summary of Contents for Aplio 500 TUS-A500
Page 41: ...No 2B771 004EN M 6 2 Systems with the main power switch on the side ...
Page 63: ...No 2B771 004EN M 9 3 Systems with the main power switch on the side ...
Page 92: ...No 2B771 004EN M 13 4 PIMS page 2 2 Refer to section 19 Storing Image Data ...
Page 163: ...No 2B771 004EN M 18 5 18 2 Needle Mark Display and Angle Change Procedures ...
Page 231: ...No 2B771 004EN M A 1 ...