8. Checks Before and After Use
In the interests of safety, it is the user's responsibility to carry out the following checks
before and after using the system.
8.1 Checks Before Turning ON the Power
Perform the following checks before the system power is turned ON and after the
system power is turned OFF.
No.
Check item
Check column
1
The temperature, humidity, and atmospheric pressure should meet the
conditions of use.
2
There should be no condensation.
3
There should be no deformation, damage, or stains to the system and
peripheral units.
* If any stains are present, perform cleaning referring to subsection 20.3.1
"Cleaning the system".
4
There should not backlash or loose screws in the casters, monitor, panel,
etc.
5
Caster locks should function correctly.
6
There should be no damage to the reference signal cable and no
looseness of the connector.
7
There should be no abnormalities such as damage, wear, or bent pins on
the power cable or plug.
8
There should be no abnormalities or stains of the transducer surface, cable
sheath, or connector, such as peeling, cracks, protruding parts, or loss of
firmness of the acoustic lens.
* If any stains are present, perform cleaning, disinfection, or sterilization
referring to the operation manual provided with the transducer.
9
No clips etc. are placed on the main panel or keyboard.
10 There should be no obstacles near the movable sections and air filter of
the system.
Summary of Contents for Aplio 500 TUS-A500
Page 41: ...No 2B771 004EN M 6 2 Systems with the main power switch on the side ...
Page 63: ...No 2B771 004EN M 9 3 Systems with the main power switch on the side ...
Page 92: ...No 2B771 004EN M 13 4 PIMS page 2 2 Refer to section 19 Storing Image Data ...
Page 163: ...No 2B771 004EN M 18 5 18 2 Needle Mark Display and Angle Change Procedures ...
Page 231: ...No 2B771 004EN M A 1 ...