8-20
8.16 Setting the LCD sleep time
[
Explanation
]
LCD is automatically turned off when there is no operation in the main body for a fixed time.
[
Operation
]
Select the Device/Other key
⇒
LCD key on the System.
①
Sleep time
Time until changing to brightness to which LCD is set by “
③
Sleep brightness” is set.
(0 to 60 min)
Please input an arbitrary “Sleep time” from the numeric input screen, and set it with the OK key.
(When 0 is input, the sleep function is not used.)
②
Act. brightness
LCD sets the numerical value of the brightness when it is active. It lightens by the numerical
value large.
Please input an arbitrary “Act. brightness (2
~
5)” from the numeric input screen, and set it with
the OK key.
③
Sleep brightness
LCD sets the numerical value of the brightness when it is sleep. It lightens by the numerical value
large.
Please input an arbitrary “Sleep brightness (0
~
4)” from the numeric input screen, and set it with
the OK key.
④
Alarm Recovery (It supports since recorder version 1.20.)
If an alarm occurs at the time of LCD OFF, LCD is turned ON automatically.
(While the alarm has occurred, regardless of [
①
Sleep time], LCD is always ON.
LCD becomes
"off" when the setup time of [
①
Sleep time] passes. (Not operating)
①
Sleep time
②
Act. brightness
③
Sleep brightness
④
Alarm Recovery