Preventative Maintenance Checklist (81000)
35
Zero-Gravity
®
Radiation Protection System Preventative Maintenance Checklist
Figure 32: ZGCMRS Shield Attachment Point
(4) 8-32 X
1.25” LG
Screw
Swivel Arm Hardware Inspection
1. While the system is not in use, inspect all hardware on the swivel arm to ensure
they are in place and fastened tightly.
2. Remove the cover on the upper arm to ensure correct hardware is in place.
3. Ensure the Phillips head screw, lock screw, and key are in place and secure
where the shield arm is inserted into the lower arm. (Figure 35)
4. Contact a TIDI Products authorized representative or TIDI Products service for
repairs and / or replacement of components. Replacement and repairs may only
be completed by TIDI Products authorized personnel.
5. If additional support is needed, contact TIDI Technical Service using the provided
contact information.
Figure 33
Slide
Cover
Upper
Arm
Lower
Arm
Cover