Preventative Maintenance Checklist (81000)
17
Zero-Gravity
®
Radiation Protection System Preventative Maintenance Checklist
Figure 7
Figure 8
Inspect for
Damage
Inspect for Cracking
Through Hinge Point
Hardware and Rotation
Points
Velcro Inspection (All Systems)
1. While the system is not in use, remove the body shield from its fixed position
inside the balancer. Lower the body shield until the top of the leaded apron is low
enough for visual inspection with the leaded apron off the ground. Use caution
when lowering the body shield to prevent damage to the leaded apron. A ladder
may be required for inspection to prevent risk of damage to the leaded apron.
2. Inspect each section of Velcro for damage, cleanliness, functionality, and
adherence to the body shield frame.
3. Replace any section of Velcro if abnormalities are observed following the
instructions provided within the Velcro replacement kit.
4. Contact a TIDI Products authorized representative or TIDI Products service for
repairs and / or replacement of components. Replacement and repairs may only
be completed by TIDI Products authorized personnel.