![TIDI Zero-Gravity Manual Download Page 25](http://html1.mh-extra.com/html/tidi/zero-gravity/zero-gravity_manual_1114370025.webp)
Preventative Maintenance Checklist (81000)
25
Zero-Gravity
®
Radiation Protection System Preventative Maintenance Checklist
Figure 18
Boom Arm to Drop Tube Hardware Inspection (ZGCM-HSA)
1. While the system is not in use, check each of the (8) 5/16-18 X 1-inch long
screws attaching the Boom Arm to the drop Tube to ensure they are in place
and torqued to 425 inch-pounds (35 foot-pounds) (48 Newton-meters) using a
calibrated inspection tool. (Figure 19)
2. Contact a TIDI Products authorized representative or TIDI Products service for
repairs and / or replacement of components. Replacement
and repairs may only
be completed by TIDI Products authorized personnel.
3. If additional support is needed, contact TIDI Technical Service using the provided
contact information.