Preventative Maintenance Checklist (81000)
19
Zero-Gravity
®
Radiation Protection System Preventative Maintenance Checklist
Figure 10
Magnetic
End Stop
Body Shield Connector Hardware Inspection (All Systems)
1. While the system is not in use, check each of the (4) body shield connector
screws to ensure they are secured and torqued to 55 inch-pounds
(4.5 foot-pounds) (6.2 Newton- meters). (Figure 11)
2. Inspect the body shield to ensure the body shield is still hanging vertically and
not at an angle (tilted – Figure 12). To adjust the body shield positioning if tilted,
loosen the (4) body shield connector screws to adjust the body shield frame
positioning. While making adjustments, make sure the body shield is secured to
prevent detachment from the body shield connector. Torque to 55 inch-pounds
(4.5 foot-pounds) (6.2 Newton- meters).
3. Contact a TIDI Products authorized representative or TIDI Products service for
repairs and / or replacement of components. Replacement and repairs may only
be completed by TIDI Products authorized personnel.
4. If additional support is needed, contact TIDI Technical Service using the provided
contact information.
Tighten screws in opposite corners (one rotation each)
until tight.
ATTENTION!
Ensure Body Shield assembly is tightly secured in
balancer connector and Body Shield is hanging vertically
(not tilted).
Figure 11