It is recommended to use a
'Circular Pupil'
by selecting the appropriate option.
Here, only the diameter in X needs to be entered and the Y diameter is set to the
same value.
Note:
The pupil diameter can be defined larger than the active camera area
defined with
Camera Image Size
in the
. So, a wavefront
measurement is enabled also in case of the beam to be analyzed is wider than
the sensor area. Please, take into account that such a calculation may suffer from
increased uncertainties because the measurement points (spots of the lenslets)
are no more equally distributed across the pupil area but are concentrated at the
center.
Pupil Center (mm)
The pupil center in mm can also be set manually or automatically.
It is recommended to enable '
Use Beam Centroid
' in order to automatically adapt
the pupil center to the measured beam centroid coordinates. In case of
fluctuations of the beam position during the measurement, the Wavefront Sensor
will automatically follow the wandered spot. If you switch to manual input you can
use the '
à
0.0
' button to set the pupil centroid to the center of the Wavefront
Sensor.
Note:
Manual definition of the pupil center requires a careful and stable beam
adjustment, otherwise the measurement results will suffer from increasing errors.
It is recommended to review the pupil size and position using the appropriate
functions in the '
Spot Field
' and '
Beam View
' displays. The '
Show Pupil'
option
must be enabled in the appropriate configuration panels.
Limit to Pupil Interior
Enabling this option will limit the wavefront calculation and display of it to the pupil
interior. Points outside the pupil, that do not go into the Zernike calculation
anyway, will not be recognized for statistical wavefront analyses.
The 3D wavefront display on the
will show only the wavefront
data within the pupil when this option is enabled.
Note:
In case the 'Measured Wavefront' is selected for display, this option
influences the total measured wavefront distortions (PV, RMS). Distortions only
within the defined pupil area are usually lower that the total measured distortions
outside the pupil.
Summary of Contents for WFS150
Page 1: ...2007 Operation Manual Thorlabs Instrumentation Optical Wavefront Sensor WFS150 WFS150C...
Page 2: ...Version Date 1 2 December 2007 2007 Thorlabs GmbH 2007 Thorlabs GmbH...
Page 6: ...General WFS150 WFS150C Manual Part I...
Page 10: ...Getting Started WFS150 WFS150C Manual Part II...
Page 25: ...Description WFS150 WFS150C Manual Part III...
Page 35: ...Operating Instruction WFS150 WFS150C Manual Part IV...
Page 103: ...Service and Maintenance WFS150 WFS150C Manual Part V...
Page 107: ...Listings WFS150 WFS150C Manual Part VI...
Page 110: ...Appendix WFS150 WFS150C Manual Part VII...
Page 116: ...114 WFS150 WFS150C Manual 2007 Thorlabs GmbH Typical Response of the CCD camera...
Page 117: ...115 Appendix 2007 Thorlabs GmbH 7 4 Wavefront Sensor Drawing...
Page 123: ......