76
WFS150 / WFS150C Manual
© <2007> ... Thorlabs GmbH
feasible. The highest number of Zernike modes must not exceed the amount of
detected spots within the pupil.
Highest order in Fourier and Optometric calculations
This control can be set to 2, 4 or 6 and forces the numerical calculation of the
Fourier coefficients M, J0 and J45 as well as the Optometric parameters Sphere,
Cylinder and Axis to consider Zernike terms up to the desired order.
Mode selection for wavefront reconstruction
This table provides a list of all calculated Zernike coefficients each listed in a row.
The number of rows is determined by the '
Fit up to Zernike Order'
control.
The check box at the left side determines whether this mode should be used for
the reconstruction of the wavefront or not. For instance, to see the reconstructed
wavefront without primitive modes piston, tip and tilt (which often dominate the 3D
graphic) use the following setting:
On the other side, in order to focus on a single or a few modes that describe the
aberration best, click on '
Select none
' and than mark the modes of interest. To
focus on third order (n = 3) aberrations, do the following selection:
Summary of Contents for WFS150
Page 1: ...2007 Operation Manual Thorlabs Instrumentation Optical Wavefront Sensor WFS150 WFS150C...
Page 2: ...Version Date 1 2 December 2007 2007 Thorlabs GmbH 2007 Thorlabs GmbH...
Page 6: ...General WFS150 WFS150C Manual Part I...
Page 10: ...Getting Started WFS150 WFS150C Manual Part II...
Page 25: ...Description WFS150 WFS150C Manual Part III...
Page 35: ...Operating Instruction WFS150 WFS150C Manual Part IV...
Page 103: ...Service and Maintenance WFS150 WFS150C Manual Part V...
Page 107: ...Listings WFS150 WFS150C Manual Part VI...
Page 110: ...Appendix WFS150 WFS150C Manual Part VII...
Page 116: ...114 WFS150 WFS150C Manual 2007 Thorlabs GmbH Typical Response of the CCD camera...
Page 117: ...115 Appendix 2007 Thorlabs GmbH 7 4 Wavefront Sensor Drawing...
Page 123: ......