By default, the graph's vertical axis is scaled automatically. In case some of the
displayed Zernike coefficients are huge compared to other distortions you like to
focus on it is advantageous to switch '
Autoscale
' off. Also in case of fluctuating
Zernike coefficients that cause a permanently changing vertical scale, it is
advantageous to disable 'Autoscale' and to define a fixed range by entering Min.
and Max. values manually.
'
Omit Tip + Tilt'
is normally switched off and all calculated Zernike
Modes are displayed. But generally, the lowest order modes Piston and the
wavefront's Tip and Tilt are of less interest. Since these lowest order modes may
have a considerable higher amplitude than the higher order modes they are
disturbing and are dispensable. Enable this option to see only Zernike modes with
index 4 and higher.
The option '
Show RMS wavefront variations in Zernike orders
' will change the
displayed data content. Not a single Zernike mode is associated to a graphical
bar but a summation of all Zernike modes of a particular order.
Summary of Contents for WFS150
Page 1: ...2007 Operation Manual Thorlabs Instrumentation Optical Wavefront Sensor WFS150 WFS150C...
Page 2: ...Version Date 1 2 December 2007 2007 Thorlabs GmbH 2007 Thorlabs GmbH...
Page 6: ...General WFS150 WFS150C Manual Part I...
Page 10: ...Getting Started WFS150 WFS150C Manual Part II...
Page 25: ...Description WFS150 WFS150C Manual Part III...
Page 35: ...Operating Instruction WFS150 WFS150C Manual Part IV...
Page 103: ...Service and Maintenance WFS150 WFS150C Manual Part V...
Page 107: ...Listings WFS150 WFS150C Manual Part VI...
Page 110: ...Appendix WFS150 WFS150C Manual Part VII...
Page 116: ...114 WFS150 WFS150C Manual 2007 Thorlabs GmbH Typical Response of the CCD camera...
Page 117: ...115 Appendix 2007 Thorlabs GmbH 7 4 Wavefront Sensor Drawing...
Page 123: ......