85
Operating Instruction
© <2007> ... Thorlabs GmbH
optical elements inserted afterwards.
Go to the tab 'Wavefront' within the
between internal and user calibration.
Note:
You will increase the accuracy of the calibration when the Wavefront
Sensor works with averages > 1, see
panel.
Note:
The generated user calibration data are only valid for the actual camera
image size, selected in the
panel. Switching to an other
camera resolution will require a new calibration!
User Calibration using a Plane Wavefront
In order to perform this user calibration you need to apply a high quality beam
with having a perfect plane wavefront. Be sure the beam is properly aligned to the
Wavefront Sensor and fills the selected active camera area completely.
Note:
The calibration fails for positions that are insufficient illuminated. Although
no error occurs the according area will be excluded in further wavefront
measurements.
Start the calibration procedure using
Menu
à
Calibration
à
Plane Wavefront
or simply click on the appropriate symbol
in the toolbar. The following panel
illustrates the required calibration setup.
Click '
Do Calibration
' to perform the user calibration and the following message
Summary of Contents for WFS150
Page 1: ...2007 Operation Manual Thorlabs Instrumentation Optical Wavefront Sensor WFS150 WFS150C...
Page 2: ...Version Date 1 2 December 2007 2007 Thorlabs GmbH 2007 Thorlabs GmbH...
Page 6: ...General WFS150 WFS150C Manual Part I...
Page 10: ...Getting Started WFS150 WFS150C Manual Part II...
Page 25: ...Description WFS150 WFS150C Manual Part III...
Page 35: ...Operating Instruction WFS150 WFS150C Manual Part IV...
Page 103: ...Service and Maintenance WFS150 WFS150C Manual Part V...
Page 107: ...Listings WFS150 WFS150C Manual Part VI...
Page 110: ...Appendix WFS150 WFS150C Manual Part VII...
Page 116: ...114 WFS150 WFS150C Manual 2007 Thorlabs GmbH Typical Response of the CCD camera...
Page 117: ...115 Appendix 2007 Thorlabs GmbH 7 4 Wavefront Sensor Drawing...