17
Running E-PAGE
™
Gels
Introduction
After loading your protein samples on the E-PAGE
™
Gels, proceed
immediately to electrophoresis using the E-Base
™
Device. Refer to page 22 for
the E-Base
™
Quick Reference Guide.
Using E-Base
™
Instructions for running E-PAGE
™
Gels on the Mother E-Base
™
Device or
Daughter E-Base
™
Device are provided below.
Note:
It is not necessary to have a gel in the Mother E-Base
™
Device if you are
using a Daughter E-Base
™
Device. However, the Mother E-Base
™
Device must
be plugged into an electrical outlet.
1.
To begin electrophoresis, press and release the pwr/prg button located on
the lower-right corner of the Mother E-Base
™
Device.
The
red light
changes to a
green light
and the digital display shows the
count down time while the run is in progress.
2.
If you are using a Daughter E-Base
™
Device, press and release the pwr/prg
button located on the lower-right corner of the Daughter E-Base
™
Device.
To add to the run time while the run is in progress, press the time button
to select the desired time and then release the time button.
Do not run an E-PAGE
™
96
Gel for more than 25 minutes or an E-PAGE
™
48
Gel for more than 30 minutes.
To interrupt or stop a run in progress, see page 21.
Mother E-Base
™
Digital
display
E-PAGE
™
96
Gel
Time button
Mother E-Base
™
Daughter E-Base
™