50 | Diagnostic
Medical Devices -Touch Controller
In presence of an off alarm or failure icon, touch the icon
and wait for failure/alarm list loading.
To stop the acoustic alarm, touch the icon
and go
back to homepage.
To permanently delete the acoustic signal, disable the
audio device (
). If other faults or
alarms occur, you will get only a visual signal.
SD card absent or damaged or not recognized.
Check SD card – see
Battery disconnected or discharged.
Table 14.
Visual Alarms List
VISUAL ALARM
MEANING
VISUAL AND
ACOUSTIC
ALARM
SOLUTION
HIGH
TEMPERATURE
Check that products have been
correctly stored in the chamber
(see
Check ambient conditions
(environment should be neither too
hot nor too cold).
Clean the condenser (see
Check that door is properly closed.
Air condition the environment.
Check the electrical system or
accidental disconnection of the plug.
LOW
TEMPERATURE
Contact Technical Support.
OPEN DOOR
ALARM
Check that door is properly closed.
Check if door has been kept opened
more than the programmed limit.
LOW
EVAPORATION
TEMPERATURE
Switch off the equipment
(see
). Disconnect the
plug from the mains, wait for some
seconds and plug it again. Switch on
the equipment and touch the running
failure/alarm icon to eliminate the
signalization.
Check that door is properly closed.
If the problem persists, contact
Technical Support.
HIGH
CONDENSING
TEMPERATURE
Switch off the equipment
(see
). Disconnect the
plug from the mains, wait for some
seconds and plug it again. Switch on
the equipment and touch the running
failure/alarm icon to eliminate the
signalization.
Air condition the environment.
Check that installation has been
properly performed (see
)
Clean the condenser (see
)
If the problem persists, contact
Technical Support.
VISUAL AND
ACOUSTIC
ALARM