Medical Devices -Touch Controller
17
| Touchscreen Controller Display Functioning
Info 1 Panel
Figure 15. INFO 1 Panel
Info 2 Panel
In Info 2 panel set parameters (setpoint, high and low temperature limits) and working equipment information are
displayed.
Figure 16. Info 2 Panel
Running or set back
alarms and failures
Working Status
N&D Function
Table 3. Working Status Strings
STRING
WORKING STATUS
PAUSE
Compressor is off, waiting for next cooling cycle.
COOLING
The compressor is ON to reach setpoint.
DEFROST
The refrigerator defrosts, warming up the evaporator.
DRIPPING
Last phase of defrosting to allow evaporator dripping.
WAITING
The compressor is ON after defrosting to re-acquire the temperature.
HEATING
Heating action active (available only for some specific models).