Medical Devices -Touch Controller
9
| Transport and Handling
Transport and Handling
The machine must be transported and handled solely in
a vertical position and following the instructions printed
on the packaging. This precaution is necessary to avoid
contamination of the compressor with oil which may
cause the rupture of valves of the cooling coils and
problems with electric motor starting.
The accessories supplied with the machine (slides, wire
shelves, drawers, baskets, etc.) are shipped inside the
unit. The machine is fixed on a wooden pallet by means
of screws, wrapped with polyethylene and packaged in
cardboard, a wooden crate or wooden case.
The machine must be handled using a forklift or pallet
truck with suitable forks (fork length at least equal to
2/3 length of the unit).
Positioning
Incorrect positioning can cause damage to the machine
and create dangerous conditions for users; therefore,
the installer must comply with the following general
rules:
Figure 1. Equipment Position
-
Set the machine in a sufficiently ventilated
environment.
-
Place the machine far from heat sources and far
from sources of electromagnetic interferences
(such as motors, generators, infrared beams,
telephones) which can have negative effects on
equipment functioning.
-
Avoid exposure to direct sunlight and air
conditioning flows.
-
Remove the supplied accessories and the wooden
pallet base.
-
Position the machine with the aid of a spirit level.
If the machine needs to be set down in
order to bring it into the installation location
it is necessary to wait at least 6 hours
before switching it on.
The manufacturer declines any
responsibility for problems due to
transport carried out in any condition
different from those specified above.
Position the machine keeping a minimum
distance of 100 mm from all walls. If the
machine is embedded in any type of
furniture, a correct air flow of the
condensing unit (compressor/fan motors)
must be always ensured. The warranty will
immediately expire, if this is not
guaranteed. (Refer to
).
For models taller than 1.5 m, installation
with wall fixing brackets is recommended.