52 | Diagnostic
Medical Devices -Touch Controller
PROTECTION
AGAINST
FREEZING
Switch off the equipment
(see
). Disconnect the
plug from the mains, wait for some
seconds and plug it again. Switch on
the equipment and touch the running
failure/alarm icon to eliminate the
signalization.
Contact Technical Support.
Refrigerator functioning is
guaranteed anyway.
S1m-S2m (only
if Dmlp Digital
Monitor is
installed)
Contact Technical Support.
Refrigerator functioning is
guaranteed anyway.
UNBALANCED
PROBES
ALARM (only if
Dmlp Digital
Monitor is
installed)
Switch off the equipment
(see
). Disconnect the
plug from the mains, wait for some
seconds and plug it again. Switch on
the equipment and touch the running
failure/alarm icon to eliminate the
signalization
Contact Technical Support.
AUXILIARY
COOLING
SYSTEM
FAILURE (only
for TWIN
models)
High temperature alarm due to long
door opening.
Contact Technical Support.
Refrigerator functioning is
guaranteed anyway.
TWIN
COMMUNICATI
ON FAILURE
(only for TWIN
models)
Switch off the equipment
(see
). Disconnect the
plug from the mains, wait for some
seconds and plug it again. Switch on
the equipment and touch the running
failure/alarm icon to eliminate the
signalization.
Contact Technical Support.
VISUAL AND
ACOUSTIC
ALARM
SOLUTION