12 | Transport and Handling
Medical Devices -Touch Controller
Placing Material Inside the
Compartment
Care must be taken when loading material to avoid
malfunctions and allow air flow to ensure temperature
uniformity inside the refrigerated compartment.
Observe the following instructions:
• Do not place material over the label indicating the
maximum permitted loading level, if present.
• Store material leaving a distance of at least 60 mm
from sides and at least 180 mm from the top of the
compartment (refer
Figure 7. Correct Material Storage
Figure 8. Incorrect Material Storage
• Do not position material in contact with or near the
).
Figure 9. Probes Position
• Do not block air vents (
Figure 10. Air Vents at the Top of the Compartment