Section 6
ThermoFlex 6-5
Thermo Scientific
Condenser Filter
Failure to clean/replace the condenser filter will cause a loss of
cooling capacit and lead to premature failure of the cooling
s stem.
hermoFlex
s
s
When it is time for a more thorough cleaning, remove the one-piece grill
and then pulling it away from the top.
he condenser framing and fins located ehind the grill assem l are
ver sharp. se caution when removing the assem l .
Note
ThermoFlex900s - 5000s water-cooled chillers have an embedded
screw(s) located at the top (and bottom) of the grill securing it to the chiller.
Loosen the screw(s) to remove the grill.
Water cooled chillers also have a fan with sharp lades ensure the
chiller is off efore removing the assem l .
For water-cooled chillers, tighten the screw(s) at the top (and bottom) of
the grill.
Replacement grill assemblies are available from Thermo Fisher.
Screw
(Water-cooled chillers)
CAUTION
- Sharp edges
Figure 6-3
ThermoFlex900s - 5000s Condenser Grill Removal
ThermoFlex3500s - 5000s water-cooled
chillers have two screws securing the
condenser. One screw on the top and one on
the bottom.
CAUTION
WARNING
WARNING
Summary of Contents for thermoflex
Page 2: ......
Page 35: ...ThermoFlex900 5000 Thermo Scientific...
Page 41: ...DI SDS 4 0 gpm 15 1 lpm ThermoFlex900 5000 Thermo Scientific...
Page 78: ...517 2014 16 2014 ThemoFlex24000...
Page 121: ...Section 3 ThermoFlex 3 21 Thermo Scientific ThermoFlex24000...
Page 156: ...Section 5 ThermoFlex Thermo Scientific Other Accessories...
Page 180: ...Section 7 7 16 ThermoFlex Thermo Scientific...
Page 186: ...Section 8 8 6 ThermoFlex Thermo Scientific...
Page 200: ......
Page 210: ......
Page 218: ......
Page 219: ......