Section 6
ThermoFlex 6-3
Thermo Scientific
particulates into the system.
Note
Before using an uid or performing maintenance where contact
with the uid is li el refer to the manufacturer s SDS for handling
precautions.
For high temperature chillers ensure the uid is at a safe temperature
elow
efore handling.
Fluid Filter Bag
Before replacing the reservoir housing ensure the reservoir sight tu e
all stopper is securel in place see next page.
On ThermoFlex900-5000s, when you remove the bag you will notice a
bag replacements turn the chiller off and remove the diffuser to inspect it
for debris/damage.
he uid velocit into the reservoir will rapidl increase with the
diffuser removed and cause splashing. urn the chiller off efore
removing the diffuser. his is especiall critical when using eth lene
or prop lene gl col.
Note
Do not operate the chiller unless the diffuser is installed.
Fluid Diffuser
CAUTION
CAUTION
WARNING
WARNING
Fluid Bag Filter
Figure 6-1
Fluid Filter Bag
Funnel Housing
When it is time to replace the bag,
gently pull up on the plastic funnel
housing to remove it and simply
pull the bag out of the chiller.
Replacement bags are available from
Summary of Contents for thermoflex
Page 2: ......
Page 35: ...ThermoFlex900 5000 Thermo Scientific...
Page 41: ...DI SDS 4 0 gpm 15 1 lpm ThermoFlex900 5000 Thermo Scientific...
Page 78: ...517 2014 16 2014 ThemoFlex24000...
Page 121: ...Section 3 ThermoFlex 3 21 Thermo Scientific ThermoFlex24000...
Page 156: ...Section 5 ThermoFlex Thermo Scientific Other Accessories...
Page 180: ...Section 7 7 16 ThermoFlex Thermo Scientific...
Page 186: ...Section 8 8 6 ThermoFlex Thermo Scientific...
Page 200: ......
Page 210: ......
Page 218: ......
Page 219: ......