Thermo Scientific Spin Tissue Processor Microm STP 120
35
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387 718 - English
other than those recommended, obtain confirmation from the
manufacturer that these methods or products are not going to
damage the unit in any way.
As an optional extra
, the unit may be supplied with a double
activated charcoal filter to neutralise any fumes given off by the
reagents used. The system controls the life of the filter in hours.
When starting the unit the following message is displayed:
The value displayed (in the example 400 hours) is the time
remaining for the filter to complete its service life. The filter should
be changed when this counter displays 0000 HOURS.
Change the filter by removing the rear cover and proceeding as
shown in the figure below.
After changing the filter the working life counter should be rest to
the original value. Access option 4.3 of the Setup Menu FILTER
RESET and press
ENTER
to reset the counter.
Changing the filter
CHANGE FILTER
0400 HOURS
Filter housing
Activated charcoal filters (2)
Cover with two (2) holding
screws