The electronic unit of the carrousel is suitably protected and
adequately shielded according to EC Standards on emission and
reception of interference. Nevertheless, the unit should not be
installed close to other high power equipment that could generate
vibrations or strong electrical interference.
The carrousel is supplied with a power cable for connection to a
standard outlet with earth. The electrical connection is located at
the rear.
CAUTION:
Before connecting the power cable, make sure that the machine is
adjusted to operate at the supply voltage.
The factory settings for the power supply are for operation with
220-240 V ~. If the unit is operated with 100 – 120 V~, open the
cover of the power plug, remove the fuse holder and invert its
position so that the visor shows 115 V the correct operating
voltage. The fuses should also be replaced by others
corresponding to rated value as indicated in the table below.
1
Insert a screwdriver here to open the cover of
the power plug and invert the fuse-holder with
the voltage selector
2
Window showing the operating voltage (in the
example 220V)
3
On/Off switch.
Pressing:
I
= start
O = stop
(the image shows the stop position)
4
Standard power plug with earth connection.
Values of the fuses
depending on the operating
voltage
110-120 V - 4 A (2 x T4AH)
220-240 V - 2 A (2 x T2AH)
Connection
&