Thermo Scientific Spin Tissue Processor Microm STP 120
33
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387 718 - English
If it is necessary to raise the basket holder or move the carrousel
during a power failure, turn the key switch located above the input
plug of the power supply cable:
n
give it a quarter turn in a clockwise direction,
n
hold the key in this position for two seconds,
n
return it to the vertical position.
The green pilot light beside the POWER indicator will flash to
indicate that the unit (except the paraffin baths) is being operated
by the emergency battery power supply.
Under these circumstances it is possible to move the basket holder
with the
€
and
•
keys to extract it should it be necessary to abort
the process.
Once the specimen holder basket is removed, the carrousel should
be lowered again with the
€
key to close the reagent vessels.
Press the
STOP
key to cancel the process that has been aborted.
After a certain period of time without any key being pressed, the
machine will turn itself off.
CAUTION
Remember that if the switch at the rear of the unit has not been
turned off, when the power supply is restored the carrousel will
start automatically (with the program aborted) and immediately
begin heating the paraffin baths.
Emergency Movements
•