Thermo Scientific Spin Tissue Processor Microm STP 120
41
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387 718 - English
This section describes the adjustments that only can be carried out
by an Authorised Technical Service Centre.
CAUTION
These values should not be modified by the operator or user
as other operating values of there processor could be
affected. These modifications should only be made by trained
technicians from an Authorised Service Centre.
Access the Technical Service Setup menu by successively
pressing the keys
STOP
,
ALARM
and
TIMER
within a maximum
interval of 1 second.
1. TEMPERATURE MENU
Options
Default
value
Comments
1.1 HYSTERESIS
0.1 to 2
0.1 °C
Establishes the connection/disconnection range
above and below the assigned temperature
1.2 MAX. TEMPERATURE
50 to 70
70 °C
Temperature limits that can be selected by the
user from the SETUP MENU
1.3 MIN. TEMPERATURE
50 to 70
50 °C
1.4 PARAF. ACTIVE
WITH TIMER
Y/N
Y
Enables keeping the paraffin baths hot during
delayed start.
2. FAN
(refer to the diagram below)
2.1 OPERATING TIME
0000 to 9999
180
Time in seconds that the fan is connected when
starting the processor
2.2 ON TIME
0000 to 9999
120
Time that the fan is operating
2.3 OFF TIME
0000 to 9999
600
Pause between fan operating periods
3. MOTOR STEP BY STEP
3.1 STIRING RATE 1
005 to 250
080
Speed in rpm for stirring 1
3.2 STIRING RATE 2
005 to 250
080
Speed in rpm for stirring 2
3.3 CENTRIFUGE SPEED
005 to 250
180
Speed en rpm for the centrifugation
4. OPTIONS
4.1 SHAKE ENABLE
Y/N
Y
Enables or disables the option of activating
vertical
SHAKE
Diagram showing fan operation
Fan:
start
on
off
T2.1 T2.3 T2.2 T2.3 Time >
Service setup
•