36
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387 718 - English
Under normal operating conditions the battery has a very long
service life.
If the alarm message "Check Battery" appears repeatedly, (for
example after a long period of inactivity) the machine should be
connected to the power supply for several hours to allow the
battery to charge itself.
If the battery does not recharge and the alarm message does not
disappear, the battery should be replaced with a new one.
Change the battery by accessing the rear part of the unit and
proceeding as shown in the figure:
Place the new battery in position and reassemble the unit by
proceeding in the reverse order.
Replacing the battery
1.
Remove the six (6)
holding screws
2.
Remove and the filter
holder and turn it face down
(be careful of the cables,
refer to
3
)
4.
Remove the two (2)
screws that hold the
battery in place
3.
Remove the connector cable
from the battery