Thermo Scientific Spin Tissue Processor Microm STP 120
17
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387 718 - English
Connection of the paraffin baths
The paraffin baths are connected from the rear of the unit.
For easier access to the rear part of the apparatus, the whole unit
can be turned on itself.
Do not turn the unit holding it by the carrousel cover. Refer to
instructions and precautions to be taken when turning the unit by
hand.
CAUTION
Before connecting or disconnecting the paraffin baths, make sure
that the unit is
disconnected from the power supply
by turning
off the main switch located at the rear of the unit.
CAUTION:
Always
use the following connections:
n
the connection marked with
the number 10
only if a third
paraffin bath has been
installed in position 10
n
the connection marked with
the number 11 only if the
paraffin bath has been
installed in position 11
n
the connection marked with
the number 12 only if the
paraffin bath has been
installed in position 12
Note the
grooves
in the edge of the processor table to allow
connection of the cables to the paraffin baths.
&
&