Thermo Scientific Spin Tissue Processor Microm STP 120
37
Microm International GmbH
part of Thermo Fisher Scientific
Otto-Hahn-Str. 1A
69190 Walldorf / Germany
387 718 - English
Should any instrument failure occur, the microprocessor displays
an alarm message on the screen and the pilot light of the
ALARM
key begins flashing.
There are two types of alarm:
n
alarms that can be resolved by the user
n
alarms that must be resolved by an authorised service
technician
Alarms that can be resolved by the user:
Alarm
Cause
Solution
SOLID PARAFFIN
The unit is attempting a vertical movement over a
paraffin bath where the paraffin has not melted
Wait until the paraffin has melted.
POWER FAILURE
The power supply to the unit has failed and
therefore the basket may have been at one
station for longer than programmed and
finalisation of the program may be delayed
Determine whether the whole process
should be repeated
PROBE FAILURE
10 1112
The unit has lost contact with the probe that
measures the temperature of the indicated
station.
Make sure that the paraffin bath is
connected to the correct connector. Refer
to 'connecting paraffin baths'
CHANGE FILTER
The working life of the filter has expired
Refer to 'Changing the filter'.
SAMPLES OUT
An attempt is being made to program a delayed
start with the carrousel not lowered.
Place the carrousel in the start position
(basket submerged in reagent vessel 1)
In Take Sample the time limit for the samples out
of the liquid has been exceeded
Lower the carrousel to submerge the
basket in the reagent vessel.
CHECK BATTERY
Low battery level has been detected on starting
the unit
Make sure that the battery is connected
and that the fuse has not blown.
Acknowledge an alarm by pressing the
ALARM
key. If the alarm
does not disappear or reproduces itself frequently, notify a service
technician.
Alarms that should be resolved by an authorised Service
Technician
HOLDER LOWERED
HOLDER IN CENTRE
HOLDER RAISED
VERTICAL FAILURE
HORIZONTAL FAILURE
HORIZONTAL MOTOR ON
REGULATOR FAILURE
HOLDER OUT OF RANGE
UNIT FAILURE
The last page of this manual includes the address of the Technical
Service Centre closest to you.
Alarms