Troubleshooting
Failure message
8-6
Sorvall RC 3BP Plus
Thermo Scientific
INTERNAL
PROGRAM ERROR
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The centrifuge detectetd an internal program fault of the
memory.
Restart the centrifuge.
If an error message appears again, turn off the
centrifuge.
The emergency lid release enables you to retrieve your
samples.
Pull out the power supply plug.
Inform Customer Service.
FREQUENCY
CONVERTER
VOLTAGE TO HIGH
The centrifuge cannot be operated.
The run does not start or the centrifuge
runs down without being braked.
The line voltage exceeds the tolerance or failure of the
braking resistance.
Restart the centrifuge.
If an error message appears again, turn off the
centrifuge.
The emergency lid release enables you to retrieve your
samples.
Pull out the power supply plug.
Inform Customer Service.
Failure message
Problem with centrifuge
Possible causes and cures