4
Operation
Rotor Temperature Equilibration
4-8
Sorvall RC 3BP Plus
Thermo Scientific
If heating, input a value that is approximately 70% of the maximum rated speed of the selected
rotor.
6. Press the
to select a continuous run.
7. Press the
and enter the desired temperature.
8. Press the
to begin temperature equilibration.
9. The centrifuge will alert you that a sample overtemperature condition exists, and that pre-cooling
is taking place. Press the
key.
10. Monitor progress of the value in the RUN TEMPERATURE display. Note the time when the
displayed RUN temperature reaches the SET temperature. From the time it does, allow the run to
continue for additional 30 minutes, then press the
key.