Please pay particular attention to the following aspects:
•
Location: well-ventilated environment, set-up on a level and rigid surface with adequate
load-bearing capacity.
•
Rotor installation: Check that the rotor is locked properly into place before operating
the centrifuge.
•
Especially when working with corrosive samples (salt solutions, acids, bases), the
accessory parts and vessel have to be cleaned carefully.
Always balance the samples.
Centrifuging hazardous substances:
•
Do not centrifuge explosive or flammable materials or substances which could react
violently with one another.
•
The centrifuge is neither inert nor protected against explosion. Never use the centrifuge
in an explosion-prone environment.
•
Do not centrifuge inflammable substances.
•
Do not centrifuge toxic or radioactive materials or any pathogenic micro-organisms
without suitable safety precautions.
When centrifuging microbiological samples from the Risk Group II (according to the
Bio-safety Manual" of the World Health Organization (WHO)), aerosol-tight biological
seals have to be used.
For materials in a higher risk group, extra safety measures have to be taken.
•
If toxins or pathogenic substances have gotten into the centrifuge or its parts,
appropriate disinfection measures have to be taken (see
Highly corrosive substances which can cause material damage and impair the mechanical
stability of the rotor, should only be centrifuged in corresponding protective tubes.
WARNING
If a hazardious situation occurs, turn off the power supply to the centrifuge
and leave the area immediately.