Operation
Switch on Centrifuge
4-2
Sorvall RC 3BP Plus
Thermo Scientific
Switch on Centrifuge
1. Turn on the power switch located on a small recessed panel in the upper right corner of the front
cabinet panel into position (select "1").
The device performs a self-check of its software.
2. Watch the control panel LEDs to be sure that all segments light.
During start-up the SET DISPLAY shows:
•
If an error is detected, an error message will appear in the SET DISPLAY.
•
If no error is found, the SET DISPLAY will display the values that was displayed when the
centrifuge was turned OFF.
•
All OPTIONS selections will also be the same as when the centrifuge was turned OFF.
Open door
1. Press the
key.
Close door
Close the door by pressing down on it lightly in the middle or on both sides of it. Two locks close the
door completely. The door will close automatically.
Note
Make sure the centrifuge is properly installed before turning it ON for the first time, (see
).
Note
If any LED segments do not light, note which ones and inform Customer Service. Continued
operation when the centrifuge is unable to display all run information correctly could mislead an
unaware observer.
WARNING
Do not reach into the crack between the door and the housing. The door is
drawn shut automatically.
Use the emergency release only for malfunctions and power failures (see
WARNING
Do not reach into the crack between the door and the housing. The door is
drawn shut automatically.
Note
The door should audibly click into place.