Service Manual, Partisol Model 2000 Air Sampler
S
ECTION
3: S
ERVICE
P
ROCEDURES
P
AGE
3-15
Revision B.000
5) Plug the 4-pin end of the Analog Input Calibration Cable into the
socket labelled “USER OUTPUT” on the back of the Partisol hub
unit.
6) Ensure that the electronics compartment door (Figure 3-1) is open.
Figure 3-14. Back of
Partisol enclosure.
Figure 3-13. Calibration
screen.
Basic
Calibration Screen
Stop
Calc
Act
Offset
Span
A/I:
1.003
0.000
0.0000
1.0000
Temp:
27.9
0.0
1.0000
Pres:
0.998
0.000
1.0000
Flow:
16.7
0.0
0.0000
1.0000
A/O:
1.00
Pot:
3.752
Flow Adj:
Hold
Edit
Hub1
Sat2
Sat3
Sat4
>
Soft Function Keys in View Data Mode
Edit
Hub1
Sat2
Sat3
Sat4
>
FTS
Pump
DecrFlw HoldFlw
IncrFlw
<
Soft Function Keys in Edit Mode
--
-
+
++
Bksp
>
+/-
<
Unshifted
Shifted
Unshifted
Shifted