![Thermo Scientific PARTISOL 2000 Service Manual Download Page 38](http://html1.mh-extra.com/html/thermo-scientific/partisol-2000/partisol-2000_service-manual_1104585038.webp)
Service Manual, Partisol Model 2000 Air Sampler
S
ECTION
3: S
ERVICE
P
ROCEDURES
P
AGE
3-10
Revision B.000
Follow these steps to check the round battery:
1) From the Partisol Main screen, press <F5:Setup> to enter the
Setup screen (Figure 3-10).
2) Press <F1:Edit> to enter the Edit Mode.
3) Change the time or date to a new value and press <Enter>.
4) Turn off the instrument, wait approximately 20 seconds and turn
the power back on.
5) From the Partisol Main screen, press <F5:Setup> to enter the
Setup screen.
6) If the time or date is the same value as entered in step 3, the
battery is functioning properly. If the time or date shows a differ-
ent value than the value entered in step 3, the battery must be
replaced. Contact R&P.
7) Restore the time or date to the correct value.
3.1.6.2.
T
ESTING
THE
B
ATTERY
-B
ACKED
RAM S
OCKET
B
ATTERY
The battery-backed RAM socket battery has an approximate 2-1/2 year storage
capacity if the unit is unpowered and stored at temperatures between 0° and 40° C.
Figure 3-10. Setup screen.
Setup
Stations:
3
Std Temp:
25.0
Run Type:
BASIC
Std Pres:
1.000
Set Flow:
16.7
Ave Temp:
99.9
A/O Type:
STAT
Ave Pres:
9.999
Curr Date:
03-Dec-99
Basic Cyc:
0
Curr Time:
16:21:36
Intv Mode:
SAMP
Edit
Calib
Diag
RS232
Audit
>
Soft Function Keys in View Data Mode
Edit
Calib
Diag
RS232
Audit
>
LCD +
LCD -
<
Soft Function Keys in Edit Mode
--
-
+
++
Bksp
>
+/-
<
Unshifted
Shifted
Unshifted
Shifted
Summary of Contents for PARTISOL 2000