Revision
B.000
Service
Manual,
Partisol
Model
2000
Air
Sampler
S
ECTION
2
: T
ROUBLESHOOTING
G
UIDES
P
AGE
2-6
2.2.1.
A
MBIENT
T
EMPERATURE
E
RROR
(T)
F
LOW
C
HART
AMBIENT TEMPERATURE ERROR TROUBLESHOOTING FLOW CHART
Error code
“T”
This error code occurs if
the ambient temperature
is out of range.
The temperature must be
between -40°C and 60°C
to be valid.
Place multimeter across test
points AMBTEMP (blue) and
AGND (black) on interface
board (see note 1).
Ensure ambient temperature
probe is properly attached to
the back of the instrument.
Voltage
nominal?
Place multimeter across
test point +15V (red) and
AGND (black) on interface
board.
No
Voltage =
~ +15V?
No
Yes
Yes
Replace ambient
temperature sensor cable
R&P P/N 51-004352-0001
Check continuity of cable
between P54 and ambient
temperature sensor back
panel connector (P52) (see
note 2).
Continuity OK?
Replace cable
R&P P/N 51-005633
No
Note:
1.The voltage at this test point is related to the
temperature according to the following
equation:
1°K = 10mVDC (nominal)
2.Pins for continuity check of cable P54 to P52:
P54
P52
Pin 1
Pin 1
Pin 2
Pin 2
-----
Pin 3
3.Pins for continuity check of cable P54 to P4:
P54
P4
Pin 1
Pin 1
Pin 2
Pin 2
Go to the statistics screen
and observe the ambient
temperature reading.
Yes
Check continuity of
cable between P4
(interface board) and
P54 (see note 3).
Yes
Continuity OK?
Replace cable
R&P P/N 51-006183
No
Replace interface/CPU
board assembly
R&P P/N 50-005496
& 10-002407
Replace interface/CPU
board assembly
R&P P/N 50-005496
& 10-002407